Diagnostics and troubleshooting
Liquistation CSF39
134
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
333
Pressure sensor
F
On
On
Medium detection not possible, sampling not
possible
• Suction line not drained before sampling
• Pressure sensor defective
1.
Check the suction line, if necessary use the
pump test under
Menu
/
Diagnostics
/
System test
/
Pump purge
.
2.
Contact the Service Department.
334
Cooling system
F
On
On
Climate control module defective
1.
Replace the climate control module.
2.
Contact the Service Department.
335
Fan defective
F
On
On
Ventilator defective
1.
Replace the ventilator.
2.
Contact the Service Department.
337
Pump hose warning
M
On
Off
End of pump hose service life will be reached
shortly
Display under
Menu
/
Diagnostics
/
Operating
time information
/
Pump tube life
1.
Schedule replacement.
2.
After replacing, reset the operating time
under
Menu
/
Diagnostics
/
Operating time
information
.
338
Pump hose alarm
M
On
Off
End of pump hose service life reached
Display under
Menu
/
Diagnostics
/
Operating
time information
/
Pump tube life
.
1.
Replace the pump hose.
2.
After replacing, reset the operating time
under
Menu
/
Diagnostics
/
Operating time
information
.
339
Liquidsensor
M
On
Off
Sensor LF1 fouled
1.
Clean the sensor soon.
2.
Set the sensitivity under:
Setup
/
General
settings
/
Sampling
/
Conductive sensor
.
340
Liquidsensor
M
On
Off
Sensor LF1 fouled
1.
Clean sensor.
2.
Set the sensitivity under:
Setup
/
General
settings
/
Sampling
/
Conductive sensor
.
343
Power supply
M
On
Off
Power supply failure
344
Program pause
C
On
Off
Sampling program paused
345
Time changeover
M
On
Off
Daylight saving time/winter time setting
Normal time (winter time) active
346
Time changeover
M
On
Off
Daylight saving time/winter time setting
Daylight saving time active
347
No sample confirm.
F
On
On
Sampling command has not been processed
1.
Check the internal cable to 1IF.
2.
Perform a software reset.
348
Read program
F
On
On
Selected program cannot be read from the
program memory
‣
Create a new program.