Operation
Liquistation CSF39
76
Hauser
Specific bottles can be assigned specific filling times with the bottle synchronization
function. For example, bottle 1 is to be filled from midnight to 2 a.m., bottle 2 from 2 a.m.
to 4 a.m. etc.. The following options are available for this:
•
None:
The sampling and bottle change times are not synchronized.
•
1. bottle change time
:
Sampling starts with the first bottle. The change to the next bottle is synchronized. For
example, a time of 2 hours was set for bottle changeover, and 00:00 was set for the
synchronization. If the program is started at 5:23 a.m., for example, bottle 1 is initially
filled. The system switches for the first time to bottle 2 at midnight (00:00), to bottle 3
at 2 a.m. etc.
•
Time of bottle number
: A specific filling time is assigned to every bottle.
E.g.: 00:00 to 02:00: bottle 1;
02:00 to 04:00: bottle 2;
04:00 to 06:00: bottle 3, etc.
If the program is started at 10:00, for example, the device starts by filling bottle 6.
It is also possible to start synchronization on a specific day of the week. For example, a
time of 24 hours was set for bottle changeover, Monday 00:00 was the time set for
synchronization, and Tuesday 8 a.m. was set for starting the program. The system fills
bottle 2 until 00:00 on Wednesday and then switches to bottle 3.
•
External signal
:
The system changes to the next bottle when an external signal is received. The external
signal first has to be configured via the binary input. The binary input can then be
selected as the source.
In the Standard and Advanced program, the bottle position is not currently restored
following a power failure.
10.3.2 Program type: Basic
With the Basic program type, you can create simple sampling programs quickly based on
time, volume and flow.
In the case of volume- and flow-controlled sampling, the inputs must be configured
appropriately beforehand. If you want to create a program and use it immediately, you
must check the configuration of the sampler before programming.
The dosing volume setting makes it possible to correctly calculate the level in the bottle
and is a reliable way of preventing the bottles from being overfilled.
Here you can adjust the bottle configuration, the bottle volume and, in the case of the
device version with vacuum pump, the correct dosing volume:
‣
Menu
/
Setup
/
General settings
/
Sampling
You can go to
Setup program
either via the overview under
Select sampling
program
or via the path
Menu
/
Setup
/
Sampling programs
Menu/Setup/Sampling programs
Function
Options
Info
Current program:
Read only
The last sampling program to be created or
used is displayed.
Status
Read only
User interface
Active
:
The sampling program has been started and
the device takes a sample as per the set
parameters.
User interface
Inactive
:
No sampling program has been started, or a
program that was running has been paused.
User interface
Pause
:
Sampling program paused.