Liquistation CSF39
Operation
Hauser
61
• Logbook name
Assign a name. Example: "01".
• Source of data
Select a data source. Example: Sensor connected to channel 1 (CH1).
• Measured value
Select the measured value to be recorded. Example: pH value.
• Scan time
Specify the time interval between two logbook entries.
• Data logbook
Activate the logbook: specify the data storage method.
2. ../
Finished
: Perform the action.
The device shows the new logbook in the list of data logbooks.
3. Select data logbook "01".
Additional display:
Log time left
.
4. Only in the case of
Fill up buffer
:
Decide to set
Overflow warning
:
On
or
Off
.
On
: The device displays a diagnostic message in the event of memory overflow.
5.
Line plotter
submenu: Specify the type of graphic representation.
10.2.5 Configuring the sampling depending on the device version
The list of functions displayed depends on the device version selected with:
• Vacuum pump
1)
• Peristaltic pump
2)
• Distribution drive
3)
• Sampling assembly:
4)
Menu/Setup/General settings/
Function
Options
Info
Sampling
Number of bottles
Choice of all possible bottle
combinations
The bottle configuration you ordered
is preset in the device.
Bottle volume
0 to 100000 ml
Factory setting
Depends on the bottle
configuration
If continuous operation is selected for
a sampling program, there is the
danger of overfilling the bottles. Do
not forget to empty the bottles!
Distribution parking
3)
Selection
• Back
• None
Factory setting
Back
Causes the distribution arm to go to
the center at the back or remain
parked in the current position when
the device is started or the program is
ended.