Commissioning
Liquistation CSF39
52
Hauser
9.3.5
Sampling programs
Difference between program types
The following box provides an overview of the differences between the Basic, Standard and
Advanced program types.
Basic (1 sampling program)
Start condition:
• Immediate
• Date/time
• Immediate activation
• Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
external signal,
• Bottle change after time or
number of samples, external
signal
• Bottle synchronization
• Multiple bottles
Stop condition:
• Program end
• Continuous operation
Manual sampling
15000 ml
MODE
1
Menu/Manual sampling
Distribution position
Bottle 1
Multiplier
A0036865-EN
1. Manual sampling is triggered by the
MAN
soft key. This pauses any program
currently running.
The current bottle configuration and the current sample volume are displayed.
You can select the distributor position. In peristaltic systems, you can also change
the sample volume.
In vacuum systems,
Multiplier
a multiple of a single manual sample can be taken
under . Specification of
Multiplier
range of adjustment 1 to 50.
2. Select
Start sampling
A new screen is displayed indicating the progress of the sampling process.
Содержание Liquistation CSF39
Страница 2: ......
Страница 188: ...www addresses endress com ...