Operation
Liquistation CSF39
62
Hauser
Menu/Setup/General settings/
Function
Options
Info
Distribution reference
Selection
• Pre sampling
• Pre bottle change
• Pre program start
Factory setting
Pre sampling
The distribution arm goes through a
reference point depending on the
option selected.
Pre sampling:
This means that the distribution arm
performs a reference test before each
individual sampling.
Pre bottle change:
This means that the distribution arm
performs a reference test in every
subprogram.
Pre program start:
This means that a single reference
test is performed before the program
starts.
Power failure
Selection
• Resume program
• Stop program
Factory setting
Resume program
Decide how the sampler should react
when it is energized after a power
failure.
Resume program:
• Time and flow-paced
The program calculates the omitted
samples and enters them in the
logbook as failed. When the
program is restarted, it continues
where it was interrupted.
• Flow-paced
No samples are entered in the
logbook during the power failure.
When the program is restarted, it
continues where it was interrupted.
Sample retries
1), 2), 3)
0 to 3
Factory setting
0
If sampling is started and no sample is
drawn in, sampling can be repeated
up to 3 times.
Sampling delay
0 to 99 s
Factory setting
0 s
The start of the sampling cycle can be
delayed by up to 99 s. The binary
output is switched without any delay.
Liquid detection
Selection
• Automatic
• Semi automatic
• Manual
Factory setting
Automatic
Automatic
The last intake time determined is the
new purge time.
Semi automatic
In
Semi automatic
purge times and
intake times can be defined
separately.
Manual
In
Manual
the dosing time can also
be set.
Dosing volume
1)
1)
20 to 350 ml
Factory setting
200 ml
4)
10 to 1000 ml
Factory setting
200 ml
1)
Adjust the dosing tube in the dosing
chamber to change the dosing
volume. The level in the bottle is
calculated using the set dosing
volume.
Conductive sensor
1)
Selection
• Low sensitivity
• Medium sensitivity
• High sensitivity
Factory setting
Medium sensitivity
The switching behavior can be set
with the liquid detection function. Use
the high sensitivity setting if the
sample has a low conductivity, for
example.
Содержание Liquistation CSF39
Страница 2: ......
Страница 188: ...www addresses endress com ...