Liquistation CSF39
Diagnostics and troubleshooting
Hauser
135
No.
Message
Factory settings
Tests or remedial action
S
D
F
349
Read program
F
On
On
Program created cannot be saved
Hardware error has occurred
‣
Contact the Service Department.
351
Delete program
F
On
On
Selected program cannot be deleted from the
program memory
‣
Perform a software reset.
352
Read Programlist
F
On
On
Program list cannot be read from the program
memory
‣
Reset the device:
Menu
/
Diagnostics
/
Device
restart
353
Overfill check
F
On
Off
Total capacity of bottle reached
No further sampling to current bottle is triggered
‣
If desired: Make changes to the sampling
program under
Select sampling program
.
354
Bottle check
F
On
Off
No empty bottles available for current program
No further sampling
‣
Check program settings under
Select
sampling program
.
355
Start time over
M
On
Off
Start time entered is in the past
‣
Enter a new start time.
356
Overfill check
F
On
Off
The total sample volume does not fit in the
sample bottle
‣
Change the sample volume.
357
Sampling faulted
M
On
Off
• Sample discarded
• There are too many sampling requests pending
‣
Make changes to the sampling program
under:
Select sampling program
.
358
Configuration
F
On
On
Program configuration does not match the current
device configuration
‣
Adjust the configuration.
359
Emptying error
F
On
On
• Error during emptying
• Emptying and sampling program is canceled
1.
Check the connection to the FMSY1
module.
2.
Check the 4R module, replace it if
necessary.
3.
Reset the device:
Menu
/
Diagnostics
/
Device restart
366
Module connection
F
On
On
No communication with the actuator module
‣
Check the internal connecting cable to the 1IF
module.
370
Internal voltage
F
On
On
Internal voltage outside the valid range
1.
Check supply voltage.
2.
Check inputs and outputs for short-
circuiting.
373
Electronictemp. high
M
On
Off
High electronics temperature
‣
Check ambient temperature and energy
consumption.
374
Sensor check
F
On
Off
No measurement signal from sensor
1.
Check sensor connection.
2.
Check the sensor, replace it if necessary.