Operation
Liquistation CSF39
70
Hauser
3. Select the desired update and select yes when asked the following:
The current firmware will be overwritten.
After this the device will reboot.
Do you want to proceed?
The firmware is loaded and the device is then started with the new firmware.
Saving the setup
Saving a setup offers the following advantages, among others :
• Copying settings for other devices
• Quick and easy switching between various setups, e. g. for different user groups or for
recurring sensor type changes
• Restoring a tried-and-tested setup, e. g. if you have changed a lot of settings and no
longer know what the original settings were
1. Insert the SD card into the controller card reader.
2. Go to
Menu
/
Setup
/
General settings
/
Extended setup
/
Data management
/
Save
setup
.
3.
Name
: Assign a file name.
4. Then select
Save
.
If you have already assigned the file name, you will be asked whether you want to
overwrite the existing setup.
5. Use
OK
to confirm or cancel and assign a new file name.
Your setup is stored on the SD card and you can upload it quickly to the device at
a later date.
Loading the setup
When you load a setup, the current configuration is overwritten.
1. Insert the SD card into the controller card reader. A setup must have been saved to
the SD card.
2. Go to
Menu
/
Setup
/
General settings
/
Extended setup
/
Data management
/
Load
setup
.
A list of all the setups on the SD card is displayed.
An error message is displayed if there is no valid setup on the card.
3. Select the desired setup.
A warning is displayed:
The current parameters will be overwritten and the device will reboot.
Warning: Please note that cleaning and controller programs can be active.
Do you want to proceed?
4. Use
OK
to confirm or cancel.
If you select
OK
to confirm, the device restarts with the desired setup.
Exporting the setup
Exporting a setup offers the following advantages, among others:
• Export in XML format with a stylesheet for formatted display in an XML-compatible
application, such as . Microsoft Internet Explorer
• Importing the data (drag and drop the XML file into a browser window)
1. Insert the SD card into the controller card reader.
2. Go to
Menu
/
Setup
/
General settings
/
Extended setup
/
Data management
/
Export
setup
.
3.
Name
: Assign a file name.
Содержание Liquistation CSF39
Страница 2: ......