Operation
Liquistation CSF39
54
Hauser
10
Operation
10.1 Display
10.1.1 Measuring mode
‣
To display the measured values, press the soft key
MEAS
in the start screen, or during
operation press
STAT
under
Measurement
.
Press the navigator button to change the mode
There are various display modes:
•
Channel overview
The names of all the channels, the sensor type connected and the current main value are
displayed.
•
Main value of the selected channel
The name of the channel, the sensor type connected and the current main value are
displayed.
•
Main value and secondary value of the selected channel
The name of the channel, the connected sensor type and the current main value and
secondary value are displayed.
Temperature sensor 1 has a special function. The states of the compressor, ventilator
and heater are displayed (on/off).
•
All the measured values of all the inputs and outputs
The current main value and secondary value as well as all the raw values are displayed.
•
User-defined measuring screens
You configure what values you want to display. You can choose from all the measured
values of physical and "virtual" sensors (calculated using mathematical functions) and
output parameters.
• Heartbeat diagnostics
Quick overview of the health of the device and of each connected sensor that supports
Heartbeat technology
In the first 3 modes, you can switch between channels by turning the navigator. In
addition to having an overview of all the channels, in the 4th mode you can also select
a value and press the navigator to see more details for the value. You can also find
your user-defined screens in this mode.
Heartbeat diagnostics
(optional or with additional activation code)
• Heartbeat diagnostics screen with graphic indicators for the health of the device and
sensor and with a maintenance or (sensor-dependent) calibration timer
• Heartbeat status information on the health of the device and the condition of the sensor
•
: Sensor/device condition or maintenance timer > 20 %; no action is required
•
: Sensor/device condition or maintenance timer > 5 ≤ 20 %, maintenance not yet
urgent but should be scheduled
•
: Sensor/device condition or maintenance timer < 5 %, maintenance is
recommended
• The Heartbeat sensor condition is the assessment of the calibration results and the
sensor diagnostic functions.
An unhappy smiley can be due to the calibration result, the measured value status or to
the operating hours limit having been exceeded. These limits can be configured in the
sensor setup in a way that adapts the Heartbeat diagnostics to the application.