Liquistation CSF39
Commissioning
Hauser
53
3. After manual sampling, a running program can be displayed and continued with the
ESC
ESC
button.
The sample volume for "Manual sampling" is not taken into account in the
calculated bottle volumes.
Programming for automatic sampling
Create a simple sampling program in the general overview under
Select sampling
program
/
New
/
Basic
or in the menu
Menu
/
Setup
/
Sampling programs
/
Setup
program
/
New
/
Basic
:
1. Enter the "Program name".
2. The settings from the
Basic setup
for bottle configuration and bottle volume are
displayed.
3.
Sampling mode
=
Time paced CTCV
is preset.
4. Enter the
Sampling interval
.
5. Enter the
Sampling volume
per sample. (For version with vacuum pump, configure
under
Menu
/
Setup
/
General settings
/
Sampling
.)
6. Select the
Bottle change mode
after number of samples or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of bottle number).
The description for this can be found in the "Bottle synchronization" section.
With the option "Bottle change after a time", you can choose the bottle
synchronization before the start condition (None, 1st bottle change time, 1st time of
bottle number). The description for this can be found in the "Bottle
synchronization" section.
1. For
Multiple bottles
enter the number of bottles the sample should be distributed
over.
2.
Start condition
: immediately or after date/time
3.
Stop condition
: after program end or continuous operation.
4. Pressing the
SAVE
saves the program and ends data entry.
Example:
15000 ml
100 ml
144
MODE
144
2
A0029242-EN
The program can be started.