4-49
Vertical Alt Main Screen
The Live Move screen prompts you to position the laser heads before starting the
live move. The defined flange is displayed graphically at the bottom of the screen.
The data acquisition angular location corresponding to the actual angular location
of the laser heads is displayed just below the flange. A graphical representation of
the data acquisition angular location is displayed in the center of the flange. The
data acquisition line starts at the 0° position, which is usually the #1 Bolt position (as
illustrated above).
Start key
Use the Start key to start the live move after the laser heads are in the chosen rota-
tional position. The Enter key can be used to do the same.
Coarse and Fine Adjustment keys
Use the Coarse or Fine Adjustment keys to move the data acquisition line clockwise
or counterclockwise around the flange by the number of degrees defined back on
the Alt Main screen. Refer to the Laser Angle Advance section on page 4-13 for
more information.
Laser Angle Advance
Use the Laser Angle Advance key on the Alt screen to define the fine and coarse
incremental adjustments to be applied to the angular position of the laser heads
during the data acquisition and live move operations. Refer to the Laser Angle
Advance section on page 4-13 for more information.
Check Lasers key
Use the Check Lasers key on the Alt screen to check the operational status of the
heads. Refer to the Check Lasers section on page 3-24 for more information.
Caution!
Do not reposition the laser heads after the live move has been started. Changing
the positions of the laser heads after starting the live move will cause the move to
be incorrect.
Содержание 2130 RBM Consultant Pro
Страница 1: ...Reference Manual 2130 RBM Consultant Pro Laser Alignment Analyzer and Fixtures Part 97191 Rev 1 ...
Страница 6: ...vi ...
Страница 20: ...1 8 Fixtures Case Layout top section ...
Страница 22: ...2 2 Setting Up and Using Laser Alignment Fixtures Laser heads front view Laser heads rear view ...
Страница 90: ...3 10 Horizontal Alignment ...
Страница 364: ...5 22 Straightness Measurements ...