4-34
Vertical Alignment
Clear key
Use the Clear key to clear the displayed data average, when the Laser Mode is set to
Averaging. When the Laser Mode is set to Standard, this key is not available.
Enter key
Use the Enter key to accept the data, after you have acquired enough data (sam-
ples). At this point the data is analyzed and a sine wave is curve fit to the data.
Note
A maximum of 20 reading sets (moves) can be stored on an align-
ment job. If more than 20 reading sets are acquired a message is dis-
played giving you the option of either discarding the last reading set
taken or overwriting the 20th reading set with the last reading.
If the fit is satisfactory (85% and above), the program will mark the Sweep Laser
Heads step back on the Main screen completed and advance you to the next step
in the procedure. In addition to marking the Sweep Laser Heads step complete,
the number of moves displayed in the upper section of the main screen is updated.
If the fit is unsatisfactory (less than 85%) a warning message displays to inform you
that the data is “Unfit”. At this point either repeat the data acquisition or try to man-
ually condition or edit the data using the edit data function. This built-in check
helps alert you to the data losing reliability in the misalignment calculations. Refer
to the Display Sine Fit section on page 3-168 for more information.
Note
Remember, when shaft movement due to causes other than misalign-
ment is present in the data this will appear as randomness (noise) in
the data. This is not normally a problem when the misalignment is
great, but will increasingly interfere as the misalignment decreases.
Therefore, the better the alignment the more likely the chance will
be of getting an unfit data warning.
Содержание 2130 RBM Consultant Pro
Страница 1: ...Reference Manual 2130 RBM Consultant Pro Laser Alignment Analyzer and Fixtures Part 97191 Rev 1 ...
Страница 6: ...vi ...
Страница 20: ...1 8 Fixtures Case Layout top section ...
Страница 22: ...2 2 Setting Up and Using Laser Alignment Fixtures Laser heads front view Laser heads rear view ...
Страница 90: ...3 10 Horizontal Alignment ...
Страница 364: ...5 22 Straightness Measurements ...