3-29
Laser Address Selection
Note
If you enter a number outside the range of 1 to 10 or try to enter two
addresses which are the same, you will get a warning telling you of the
error. Should this happen, press the Enter key to continue and repeat
the procedure.
Read Head B Address key
Use the Read Head B Address key to read and set the address in Head B. The pro-
cedure for Head B is the same as described above for Head A, except in this case
Head B is turned on and Head A is turned off. If a new address is entered it must
be different than the Head A address.
Set Analyzer Address key
Use the Set Analyzer Address key to read and set the address in the analyzer.
The
analyzer address can be set from 11 (default) to 15.
Note
If you enter a number outside the range of 11 to 15 you will get a
warning telling you of the error. If this occurs, press the Enter key to
continue and repeat the procedure.