Figure F 2. Replace the dust filter foam
3. Transportation or Return to Factory
3.1 Instrument Disinfection
In case the Gentier instrument will be moved to another lab or needed to return to the after-sale
service department for maintenance, please first disinfection the instrument and fill in the
disinfection certificate.
The disinfection process of Gentier instrument is listed as follows:
▶
Please wear protective clothing and medical disposable gloves;
▶
Open the loading platform and get all consumables out of the sample block;
▶
Switch offthe instrument and unplug the power cord;
▶
Prepare the detergent (lab routine disinfectant), wet the cotton cloth and degreased cotton
swabs with the prepared detergent;
▶
Clean the loading platform with the cotton cloth, clean the sample block well with the
degreased cotton swabs, and leave the loading platform open until it is dry;
▶
Clean the loading platform and the sample block again with 75% ethanol;
▶
Clean the instrument shell with mild detergent;
▶
Open the loading platform and leave the instrument under UV light for 1 hour disinfection.