Figure E 3.7 Temperature Setting sub tab Step Edit window Gradient step mode
The gradient step mode allows the Gentier sample to adopt different temperatures. After
setting the temperature center value and the temperature offset value for the current
gradient step, the system will automatically calculate the gradient temperatures.
User can press the input boxes of
Temp. Center
and
Temp. Offset
to edit the
corresponding parameters for the current gradient step. User could also and press
<
Details >
to view the specific gradient temperatures of each column, as shown in figure
E 3.7a.
Figure E 3.7a Step Edit window Gradient step parameter setting
Temp. Center
: the center value of the temperature grading range.
Temp. Offset
: the offset value of the temperature grading range.