Gentier Real-Time PCR System
3.2 External Device Connection
Please plug the corresponding power cords of Gentier instrument, control computer and printer
to the power supply respectively. Then please confirm the connections of external devices, such
as the control computer, display, keypad, mouse and printer, etc. The external device
connnection is completed.
3.3 Application Software Installation
The application system software is required on the control computer for commanding the
Gentier instrument and analyzing the experimental data. Before starting up the application
software, user should install it on the control computer. The specific operations are as follow:
1
st
step
: switch on the control computer and close its firewall.
2
nd
step
: place the Gentier software CD in computer’s CD drive.
3
rd
step
: the application software installation interface will automatically pop up, please follow
the prompt to install the software.
4
th
step
: double-click the MED-TL
icon on the desktop to startup the Gentier aplication
software.
3.4 Computer Network Setting
The network address of the control computer should be set according to the default network
information of the Gentier instrument. So the control computer can connect the identifiable
Gentier instruments within LAN.
1
st
step
:
2
nd
step
:
3
rd
step
:
please first check the network information of Gentier instrument, which is
displayed on the network information window of the instrument software, as
shownin figure B-4a. For detiled operation, please refer to
E-2.4 Gental Settings –
2.4.2 Configuration Sub-tab – NetworkInformation
.
please open the
Contol Panel > Network and Sharing Center > Loacl Area
Connection > Peroperties > Internet Protocal Version 4 ( TCP/IPv4)
on the
control computer, as shownin figure B-4b.
set the IP address, subnet mask and default gateway of control computer according
to the default network information of the Gentier instrument .