78
Gentier Real-Time PCR System
Figure D-13. Melting Curve analysis interface
Interface Introduction and Parameter Description
Reminding
: the parameter descriptions of
Sample Setting, Result Table
and
Heat Map
functional modules are same with those introduced in
D-3.3.4.1 Abs Quant
.
I.
Melting Curve
: display the fluorescence intensity against the temperature, as shown in
figure D-13.1.
▶
▶
Figure D-13.1 Melting Curve analysis interface - Melting Curve functional module
Parameter introduction:
X axis represents the temperature value;
Y axis represents the Relative Fluorescence Unit (RFU);
Function key introduction
:
User can select the melting curve display mode from the drop-down list on the bottom left
of
Melting Curve
functional module. The melting curve can be displayed according to
Well
Color
,
Dye Color, Sample Color
and
Gene Color
.
The introduction of
<
Y-axis Coordinate Adjustment >
and
<
Export >
keys are
same with those keys introduced in
D-3.3.4.1 Abs Quant - Amplification Curve
.
Reminding
: user can hover the mouse over a certain melting curve, the curve will be
highlighted and its corresponding channel and well coordinate will be displayed on the
right of cursor.
Reminding
: user can drag and drop the left mouse button to zoom in the local melting