76
Gentier Real-Time PCR System
analysis of the current experiment within the gene and sample tab, as shown in figure D-12.
Figure D 12 Rel Quant Analysis Setting interface Gene and Sample tab
▶
Dye and Gene
: display the name oftarget gene and its marking dye.
a.
User can check a certain check box in
RefGene
column and click
< OK >
to set the reference
gene of the current experiment.
b.
User can check a certain check box in
Remove
column and click
< OK >
to remove the
experiment data of corresponding dye labeled genes from the Analysis interface.
c.
User can click a certain
Eficiency
column, use
and
keys to set the amplification
efficiency ofthe corresponding dye labeled gene.
Reminding
: if a standard curve has obtained for the current experiment, the application
software will automatically load the amplification efficiency of the standard curve.
▶
Sample
: display all the sample name of sample setting.
a.
User can check a certain check box in
Ref Sample
column and click
< OK >
to set the
reference sample of the current experiment.
b.
User can check a certain check box in
Remove
column and click
< OK >
to remove the
experiment data of samples with corresponding sample name from the relative