102
Gentier Real-Time PCR System
Figure E-3.3 Cycle edit box
▶
User could set the cycle number in the cycle edit box, press
< OK >
to confirm the cycle
setting or press
< Cancel >
to cancel the setting.
Reminding
: the cycle setting range is 1~99.
Step 3
:
Add/delete a stage
▶
Select a stage in the stage setting box and the step setting box on the right will display the
corresponding steps of the selected stage;
▶
User can add a new step or select a step to delete or change its order;
Reminding
: at least one step should be included in one stage.
Step 4
:
Set fluorescence reading step
▶
In the step setting box, user can press the
Fluor
coloum of a cetain step to set whether to
read its fluorescence, the
icon will be displayed in the corresponding
Fluor
coloum of
the fluorescence reading step.
Reminding
: user can set only one fluorescence reading step in one stage.
Step 5
:
Edit step
User can select a certain step in the step setting box, and press the
icon in the
corresponding
Edit
column, the instrument software will automatically pop up the step edit
window, as shown in figure E-3.4.