Gentier Real-Time PCR System
Ct
: user can check
Ct
option and click any dye to view the relevant Ct heat map.
Concentration
: user can select
Concentration
option and click any dye to view the relevant
concentration heat map.
Fluorescence
: user can select
Fluorescence
option and select the cycle number from drop-
down list; then click any dye to view the the relevant fluorescence heat map at certain cycle.
QC
: user can select
QC
option, and if a certain sample or reference conforms to the QC
failure condition set on the Sample Setting interface, then the relevant sample well will be
displayed as
N/A
on QC diagram.
Abs Quant Analysis Setting
User can click the
<
Analysis Setting >
icon in the Tool Bar and the application software
will automatically pop up the analysis setting window, which is composed of
Amplification Plot
and
Gene and Sample
two tabs. User can set corresponding parameters for the current
experimental data analysis method in the analysis setting window.
I.
Amplification Plot
tab: user can set the
Analysis Mode
and
Analysis Method
of the current
experiment within the
Amplification Plot
tab, as shown in figure D-10.1.
Figure D-10.1 Abs Quant Analysis Setting interface - Amplification Plot tab