123
Gentier Real-Time PCR System
2.4 Replace Fuse Tube
The fuse tube (type F10AH250V) of Gentier instrument is located in the fuse tube box near the
power outlet at rear of the instrument. Before replacing the fuse tube, please switch off the
instrument and unplug the power plug, use a screwdriver to pry open the fuse tube box. Then
substitute the old fuse tube with an equal one, as shown in figure F-1.
Figure F-1. Replace the fuse tube
Reminding
: in case there is no display on the screen after starting up the Gentier
instrument, please check the fuse tube
.
Caution:
improper fuse tube may lead to circuit system damage, even fire.
2.5 Replace Dust Filter Foam
In order to avoid the dust particles from entering the Gentier instrument, every cooling fan the
instrument is equipped with dust-proof net. Please replace the dust filter foam within the
dust-proof net on a regular schedule (every six months), as shown in figure F-2. In case there
are dusts accumulated on the cooling fan, please clean it with a vacuum cleaner.