For protection against overheating hazards, the openings on the Gentier instrument are
designed for ventilation. Please do not block or cover these ventilation openings while it is
running.
7)
Do not place the Gentier instrument on soft surface, the instrument base may sink into the
soft surface and block up the air inlet beneath.
8)
Do not place the Gentier instrument close to the wall or pile up other goods at the rear of the
instrument, this may affect heat dissipation.
2.2 Instrument Power Requirements
1)
The power grid environment of Gentier instrument should possess ground wire and the
instrument should be properly grounded.
2)
Please ensure the power strip of Gentier instrument has 3~4 three-phase plugs in order to
meet the demands for the instrument, control computer and printer.
3)
The power specifications of Gentier instrument are listed in
“A3.1 Instrument Parameters
and Characters - 3.1.1 Instrument General Parameters”
, use improper power may
damage the circuit system and cause fire, it is recommended to use UPS power supply.
Reminding
: before connecting the AC power supply, please ensure the consistent of the
Gentier instrument required voltage and the power supply voltage (allowable deviation ±
10%). And make sure that the rated load of receptacle is no less than the requirement of
the Gentier instrument.
Reminding
: if the power supply system of the Gentier instrument working environment is
unstable, please do not connect other electrical equipment at the same power circuit, and
it is required to equip with over 900VA AC regulated power supply.
Prohibit
: spraying liquid on electrical parts may cause a short circuit and result in fire, do
not use sprays in vicinity of the instrument.
Caution
: do not put anything on the power cord and keep it away from places where
people move around. Hold the power plug when you plug the power cord and make sure
the power plug is perfectly inserted into the socket, do not pull the power cord to pull out
the plug.
Reminding
: under normal circumstances, please use the instrument attached power cord.