43
Gentier Real-Time PCR System
User can click
<
Up >
icon to move up the selected step.
User can click
<
Down >
icon to move down the selected step.
Reminding
: user can also click
<
Insert Step >
or
<
Delete Step >
above any step
box in the temperature program area to add or delete a temperature step.
5. Step Setting
: double click any step in the step list or in temperature program area to edit
the selected step, the application software wiil pop up the
Step Setting
dialog box for user
to edit the selected step, as shown in figure D-5.6a.
Figure D 5.6a Step Setting dialog box Standard step mode
Introduction to Step Setting Dialog box
Step Mode
: user can set the step mode for the selected step in the drop-down list of
Step Mode
,
which including
Standard, Touchdown, Long
and
Gradient
.
Model
Step Mode
Gentier 96E
Gentier 96S
Gentier 96R
Gentier 96C
Standard
Touchdown
Long
Gradient
I.
Standard Step Mode
: user can set the
Temperature, Time, Ramp
for the standard step and
decide whether to read its
Fluorescence
, as shown in figure D-5.6a.
▶
Fluorescence
: user can decide whether to read the fluorescence at current step from the
drop-down list of
Fluorescence
according to the experiment requirements.