Gentier Real-Time PCR System
3)
Figure E-3.5a Step Edit window - Touchdown step parameter setting
Initial Temp.
: the initial value of annealing temperature change range.
Target Temp.
: the target value of annealing temperature change range.
Delta Temp.
: the temperature change(°C) per cycle.
Start Cycle
: the cycle number after which the temperature change is started.
Reminding
: the initial and target temperature setting range is 35.0°C ~100.0°C.
Reminding
: the delta temperature setting range is 0.1°C ~5.0°C.
Reminding
: the star cycle range is 1~ max cycle number ofthe current stage.
Reminding
: the other parameter settings are same with the standard step mode.
If the
Long
step mode option is selected, the step edit window will be shown as E-3.6.
Figure E-3.6 Step Edit window - Long step mode
▶
The long step mode allows the temperature program to change the elongation step