10
Gentier Real-Time PCR System
Reminding
: please check the package integrity before open it. In case there is any damage,
collision or water mark, please contact the transport department or our company.
Reminding
: please check the instrument and accessories according to the packing list and
ensure that all components are present and intact, report any missing items to
XATL Co.,
LTD.
Reminding:
please fill in the relevant content on installation information feedback table
after acceptance, and send it back to our company for document and warranty.
Caution:
please reserve the original packing materials for future use. The transport
package of Gentier instrument is designed to reduce the instrument damage and ensure its
transportation safety. Adopt other packaging materials will break the warranty, and
XATL
Co., LTD.
will not be responsible for damages as consequences of improper packaging that
incurred during the transportation. Please also keep instrument related documents
provided by our company for future use too.
2. Working Environmental Requirements
The Gentier instrument is for indoor use only.
2.1 Instrument Space Requirements
1)
The Gentier instrument should be placed in the room with low humidity (between 20%
~85%RH) and appropriate temperature (between 10°C~30°C). The room should be well
ventilated and without corrosive gas.
2)
The Gentier instrument should be placed on steady lab work bench and leave enough space
for placing the attachments, cables and reagent bottles, etc.
3)
Please keep the Gentier instrument away from heat sources (directly sunshine, heater,
stoves) and water sources (such as water pool, water tube etc.).
4)
The working environment of Gentier instrument should be without electromagnetic
interference, vibration and high frequency wave electrical equipment.
5)
For single Gentier instrument running, the distance between adjacent objects and the
surrounding ventilation openings shall be no less than 30cm. In case that multiple
instruments are running at same time, the distance between adjacent instruments should
not be less than 80cm.