Gentier Real-Time PCR System
Reminding
: the temperature increment setting range is 0.1°C - 5.0°C.
6)
If the current step is the last step of a continuous melting stage, the step setting window of
this step is as shown in figure E-3.9.
Figure E-3.9 Step Edit window - Continuous Melting step mode
▶
Continuous melting stage allows the instrument system to read fluorescence more
frequently, user can press the input boxes of
Readings
to set the fluorescence reading times
per °C for the current continuous melting step, as shown in figure E-3.9a.
Readings
: the fluorescence reading times per °C.
Reminding
: the reading frequence setting range is 2readings/°C - 15readings/°C.
Figure E-3.9a Step Edit window - Continuous Melting step parameter setting