Instructions for Use
10 Troubleshooting
OPMI LUMERA 700
10.3 Faults (without messages)
G-30-1673-en - 14.0 - 2018-07-12
189 / 236
Fault
Cause
Remedy
Functions are being triggered
unintentionally.
A button on the 14-function
foot control panel has become
mechanically stuck after being
pressed.
u
Place the 14-function foot
control panel in its rest position.
u
Reconfigure the button
functions. This can only be done
if the buttons on the device are
configurable.
The 14-function foot control
panel is sending an erroneous
activation signal.
u
Place the 14-function foot
control panel in its rest position.
u
Reconfigure the button
functions. This can only be done
if the buttons on the device are
configurable.
The wrong 14-function foot
control panel is being used.
u
Check the marking. The marking
on the stand and the indicator
on the 14-function foot control
panel must match.
u
10.3.4 Wired 14-function foot control panel faults
Fault
Cause
Remedy
The 14-function foot control panel is
not functioning.
The 14-function foot control
panel is not plugged in.
u
Check to make sure the 14-
function foot control panel is
properly plugged in on the
stand.
Intermittent loss of functionality.
Failure of individual button
functions.
u
Reconfigure the button
functions. This can only be done
if the buttons on the device are
configurable.
The surgery profile cannot be
changed via the configured button
on the 14-function foot control
panel.
You are not in the main menu.
u
Return to the main menu.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...