91 / 236
• The IT network to which the device is to be connected must
have the following features:
– IPv4, static or dynamic address.
– Any IP address.
• The IT network to which the device is to be connected must
have the following configuration:
– The firewall is open for inbound ports.
– All data is transmitted unsecured from the device to the
customer's secured internal network.
– No data is transmitted externally from the device.
• Requirements in relation to network connection with the
device:
– The IT network has a RJ45 network connection, at least
Cat.5, 100 Mbit/s.
– The device has a network isolator in compliance with IEC
60601-1 which means it is not connected to the IT network
power supply.
– The operator is responsible for the topology of the IT
network, the flow of information and the routing between
the device, the IT network and other devices.