4 Installation
Instructions for Use
OPMI LUMERA 700
4.3 Connections
96 / 236
G-30-1673-en - 14.0 - 2018-07-12
4.3.3.2 Setting up a direct network connection (without IT network
connection)
1
2
Figure 59: Network connection with CALLISTO eye Panel PC (model I)
1
2
Figure 60: Network connection with CALLISTO eye Panel PC (model II)
1
LAN 1 network connection
2
LAN 1 network connection
Material
• Network cable, 10m
The network connection is established during startup. Therefore
observe the following steps:
Action
1. Turn the device and the CALLISTO eye Panel PC off.
2. Disconnect the CALLISTO eye Panel PC from the power.
3. Connect the network connections of both devices. Cable:
network cable, 10 m
4. Reconnect the CALLISTO eye Panel PC to the power.