11 Technical specifications
Instructions for Use
OPMI LUMERA 700
11.9 Guidance and manufacturer´s declaration – electromagnetic immunity
214 / 236
G-30-1673-en - 14.0 - 2018-07-12
Magnetic field for supply
frequency (50/60 Hz) as per IEC
61000-4-8
30 A/m
30 A/m
Voltage dips, short interruptions
and voltage variations as per IEC
61000-4-11
0% U
T
for 1/2 cycle
0% U
T
for 1/2 cycle
0% U
T
for 1 cycle
0% U
T
for 1 cycle
70% U
T
for 25/30 cycles
70% U
T
for 25/30 cycles
0% U
T
for 250/300 cycles
0% U
T
for 250/300 cycles
11.9.1.3 Electromagnetic immunity for non-life-supporting ME
equipment and ME systems
The OPMI LUMERA 700 is intended for operation in an electromagnetic environment as specified below. The
customer or the user of the OPMI LUMERA 700 is responsible for ensuring that the device is operated in such
an environment.
Electromagnetic immunity tests
IEC 60601 test level
Compliance level
Conducted HF disturbances as per
IEC 61000-4-6
3 V
150 kHz to 80 MHz
3 V
6 V
ISM and amateur radio bands
between 150 kHz and 80 MHz
6 V
Radiated HF disturbances as per
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
Radiated HF disturbances from
near fields of wireless communi-
cation devices as per EN
61000-4-3
27 V/m
385 MHz
27 V/m
28 V/m
450 MHz, 810 MHz – 2.45 GHz
28 V/m
9 V/m
710 MHz – 780 MHz,
5.24 GHz – 5.785 GHz
9 V/m
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...