4 Installation
Instructions for Use
OPMI LUMERA 700
4.3 Connections
104 / 236
G-30-1673-en - 14.0 - 2018-07-12
4.3.7 Connecting the strain relief
The strain relief protects the cables against being pulled out.
Action
1. Form a loop with the cable.
Length of cable between strain relief and connector: at least
320 mm
2. Open the flap.
3. Push the cable through the opening.
4. Close the flap.
5. Tighten the cable until the flap is enclosed.
6. Check the remaining length of cable between the strain relief
and the connector.