5 Daily startup
Instructions for Use
OPMI LUMERA 700
5.1 Outfitting the surgical and assistant's microscopes with accessories
112 / 236
G-30-1673-en - 14.0 - 2018-07-12
2. Hold the suspension arm in position and gently pull the [Fixing
suspension arm position] rotary knob out. If the rotary knob
does not release, repeat step 1.
3. Release the magnetic brakes. To do this, press and hold the
[Release magnetic brakes] button or the button configured for
this purpose on the surgical microscope handgrip.
4. Move the suspension arm up and down approximately 20 cm
while adjusting the [Weight balancing] rotary knob until the
force of the upward movement is exactly the same as that of
the downward movement.
Result
ü
If the surgical microscope remains in position with the
magnetic brakes released, the weight was successfully
balanced.