Parameter
Symbol
Effect
Large spot
The exposure is measured in an
area in the image center. This light
metering method is ideal for
surgical fields with darkened
edges.
Small spot
The exposure is measured in a very
small area in the image center.
This light metering method is
suitable for working with an
extremely small illuminated field
diameter. If the object you are
interested in viewing is not located
in the center of the image, set the
light metering method to "Large
spot" or "Integral".
Prerequisite
þ
The device includes an integrated HD camera.
Action
1. Open the main menu.
2. Tap the [Camera] menu button.
3. Tap [Tab 2].
ð
The "Camera" menu will appear.
4. To set the automatic exposure: deactivate the [Manual
brightness] button.
ð
The light metering methods will be activated, and
"Brightness" will be written in the left values range.
5. Select the desired light metering method.
6. Under "Brightness", set the desired brightness.
7. To temporarily save the settings: tap the
button.
8. To save the settings to the surgery profile: tap the
button.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...