9 Maintenance
Instructions for Use
OPMI LUMERA 700
9.3 Performing safety inspections
182 / 236
G-30-1673-en - 14.0 - 2018-07-12
CAUTION!
Risk of injury without safety inspection!
Hazards and device deficiencies will not be detected in time and
can have a negative effect on patients, users, or others.
u
Have a safety inspection in compliance with IEC 62353
performed on time and to the prescribed extent. At the same
time, be sure to observe the relevant national regulations.
Prerequisite
þ
Only the manufacturer or qualified persons may perform safety
inspections.
Action
u
Check to make sure the Instructions for Use are available.
u
Check to make sure all marks and inscriptions on the device are
legible.
u
Check the leakage current and the protective ground
conductors.
u
Check to make sure the casters and locking devices function
properly and are not worn-out.
u
Perform a function check of all switches, buttons, sockets and
indicator lamps of the system.
u
Check every six months to make sure manual mode is working.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...