7.5.8.3 Setting manual exposure control
Manual exposure deactivates the exposure metering and sets a
fixed exposure time. The exposure time can be set to any value
between 1/10000 s and 1/8 s. An exposure time in the range
between 1/50 and 1/30 usually provides optimum exposure results.
Prerequisite
þ
The device includes an integrated HD camera.
Action
1. Open the main menu.
2. Tap the [Camera] menu button.
3. Tap [Tab 2].
ð
The "Camera" menu will appear.
4. To set the manual exposure: activate the [Manual brightness]
button.
ð
The light metering methods will be deactivated, and
"Exposure time" will be written in the left values range.
5. Under "Exposure time", set the desired exposure time.
6. To temporarily save the settings: tap the
button.
7. To save the settings to the surgery profile: tap the
button.
7.5.8.4 Configuring color values
Prerequisite
þ
The device includes an integrated HD camera.
þ
A white balance was performed. [
Action
1. Open the main menu.
2. Tap the [Camera] menu button.
ð
The "Camera" menu will appear.
3. Tap [Tab 2].
4. Use the arrow buttons to change the color values for Hue and
Chroma.
5. Tap [Tab 3].
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...