6 Before every use
Instructions for Use
OPMI LUMERA 700
6.2 Switching the device on
124 / 236
G-30-1673-en - 14.0 - 2018-07-12
6.2 Switching the device on
1
2
Figure 72: Switching the device on (example with integrated CALLISTO eye
Panel PC)
1
[Power] button
2
Power switch
CAUTION!
Risk of infection caused by particles being stirred up!
The built-in fan can stir up loose particles and transfer germs to the
patient when the device is switched on.
u
Only turn the device on when there is no patient beneath the
surgical microscope.
NOTE
Device cannot be operated!
The automatic circuit breaker in the power switch was activated
due to a circuit overload or short.
u
To switch the device on: press the power switch.
u
If the automatic circuit breaker is triggered again, notify ZEISS
Service.
Prerequisite
þ
The 14-function foot control panel is connected.
þ
For a direct network connection between the device and the
CALLISTO eye Panel PC, both devices must be connected by
means of a network cable.
Action
1. Press the power switch.
ð
The power switch will illuminate green and the main menu
will appear on the control panel.
ð
The device is ready to be started.
2. If the CALLISTO eye Panel PC is connected: press the [Power]
button.
ð
The [Power] button will illuminate green.
ð
CALLISTO eye is ready for use.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...