Instructions for Use
5 Daily startup
OPMI LUMERA 700
5.2 Adjusting the surgical and assistant's microscopes
G-30-1673-en - 14.0 - 2018-07-12
113 / 236
5.2 Adjusting the surgical and assistant's
microscopes
5.2.1 Adjusting the downward travel limit
The downward travel limit limits the working distance from the
surgical field. This prevents the surgical microscope from coming
into contact with the patient if it is inadvertently lowered.
Action
1. Release the downward travel limit lock. To do this, rotate the
[Downward travel limit] rotary knob counter clockwise.
2. Lower the surgical microscope until you are able to focus while
maintaining sufficient working distance from the surgical field.
To do this, press and hold the [Release magnetic brakes] button
or the button configured for this purpose on the surgical
microscope handgrip.
3. Lock the downward travel limit. To do this, rotate the
[Downward travel limit] rotary knob clockwise as far as it will
go.
ð
The downward travel limit is set.
4. Check the working distance from the surgical field. To do this,
lower the surgical microscope to its lowermost stopping
position again.