10 Troubleshooting
Instructions for Use
OPMI LUMERA 700
10.4 Troubleshooting work
192 / 236
G-30-1673-en - 14.0 - 2018-07-12
10.4.3 Changing the halogen lamp
1
2
4
5
3
Figure 79: Changing the halogen lamp
1
[Open lamp module] button
2
Lamp module
3
Lamp reflector of the
halogen lamp
4
Lamp bulb of the halogen
lamp
5
Ceramic base of the lamp module
CAUTION!
Risk of burn injury caused by a defective halogen lamp!
Damage to new halogen lamp caused by body oils.
u
Wait several minutes until the halogen lamp is cool.
u
Replace the halogen lamp with heat-reflective protective
gloves.
Material
• Halogen lamp: 12 V, 100 W
Prerequisite
þ
The power switch is off.
Action
1. Press the [Open lamp module] button.
ð
The lamp module will eject partially.
2. Pull out the lamp module.
3.
CAUTION! Risk of burns due to hot halogen lamp!
Remove
the defective halogen lamp from the spring-loaded mount.
4. Pull the ceramic base from the contact pins of the halogen
lamp.
5.
NOTE! New halogen lamp can be damaged by contact!
Plug the ceramic base onto the contact pins of the new
halogen lamp.
6. Insert new halogen lamp.
7. Press the halogen lamp into the spring-loaded mount.
8. Push the lamp module, including the new halogen lamp, back
into the device.
9. Turn on the device at the power switch.
10. Check both halogen lamps for functionality.
Summary of Contents for opmi lumera 700
Page 1: ...OPMI LUMERA 700 Software Release 3 1 Instructions for Use...
Page 10: ...Empty page for your notes...
Page 14: ...Empty page for your notes...
Page 30: ...Empty page for your notes...
Page 130: ...Empty page for your notes...
Page 200: ...Empty page for your notes...
Page 222: ...Empty page for your notes...
Page 226: ...Empty page for your notes...
Page 230: ...Empty page for your notes...
Page 235: ...Empty page for your notes...