background image

 

 

 

 

 

 
 
 
 

 

 

 

 

 

 

73 Mall Drive, Commack, NY  11725  •  631-543-2000 (P)  •  631-589-6975 (F) 

www.unitronusa.com  •  [email protected] 

RMM2 

 

ROLLSCOPE MICROSCOPE 

 
 

MANUAL 

 

Summary of Contents for RMM2

Page 1: ...73 Mall Drive Commack NY 11725 631 543 2000 P 631 589 6975 F www unitronusa com info unitronusa com RMM2 ROLLSCOPE MICROSCOPE MANUAL ...

Page 2: ... 631 543 2000 www unitronusa com 2 CONTENTS SAFETY NOTES 3 CARE AND MAINTENANCE 3 INTRODUCTION 4 UNPACKING AND COMPONENTS 4 COMPONENTS DIAGRAM 6 ASSEMBLY 7 OPERATION 7 PHOTOMICROGRAPHY 10 ADJUSTMENT AND CARE 11 SPECIFICATIONS 12 MAINTENANCE 13 SERVICE 13 WARRANTY 13 ...

Page 3: ...OTE Always plug the stereomicroscope power cord into a suitable grounded electrical outlet A grounded 3 wire cord is provided CARE AND MAINTENANCE 1 Do not attempt to disassemble any component including eyepieces objectives or focusing assembly 2 Keep the instrument clean remove dirt and debris regularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt sh...

Page 4: ...power objectives as well as in choosing and composing areas for photography UNPACKING AND COMPONENTS Your microscope arrived packed in a molded shipping carton Do not discard the carton the shipping carton should be retained for reshipment of your microscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form Carefully remo...

Page 5: ...t photographs are being made PLANE GLASS REFLECTOR No adjustment is necessary to provide ideal conditions for vertical illumination The plane glass reflector has been factory installed adjusted and inspected and is correctly positioned to assure even full field illumination No positioning change is required before use unless the reflector has been disturbed in transit assembly or usage DIAL DEPTH ...

Page 6: ...icrometrics Digital Camera Depth Gauge Depth Gauge Bracket With Lock Screw X Axis Control Graduated Positioning Scale Fine Focus Knob Roll Feet Contact Screw Vertical Illuminator Not shown Bulb cord and socket assembly Bulb Centering Screws Y Axis Control Coarse Focus Knob Field Diaphragm Control Lever Illuminator Filter Slot Objectives Nosepiece Optional Accessory ...

Page 7: ...life In general use higher illumination intensities to provide short exposure times during photography or for examining poorly reflecting materials Always turn the transformer off when the microscope is not in use Operation With Batteries The illuminator may be operated by batteries for completely portable operation NOTE Insert the metal battery spacer into each foot before inserting the batteries...

Page 8: ... tightened or loosened The bulb position is optimum when the filament image appears as shown in Figure C below Figures A and B illustrate an off centered filament Lamp Filament Alignment 2 Referring to the figure above note that there are three conditions to be met a The full square area of the filament should be seen This is obtained by rotating the bulb socket in the sleeve of the centering adju...

Page 9: ...imen is in sharp focus A slight adjustment with the fine focus knob may be necessary when changing objective power Changing Objective Magnifications Total magnification is the product of the eyepiece power multiplied by the power of the objective The objectives are parfocal Therefore once an image is in sharp focus only a slight adjustment of the fine focus control will be needed when rotating the...

Page 10: ...auge reading gives a direct measurement of the difference in height between the two levels The highest measuring accuracy will be obtained by using the 50x objective since this objective has the smallest depth focus NOTE This procedure is not necessary with the digital depth gauge X Y Motion The RMM2 Series is equipped with a built in mechanical bench which slides the microscope body in X Y direct...

Page 11: ...t cover The eyepiece should always be in the microscope to eliminate the possibility of dirt or dust from entering the vertical tube and falling onto the back lenses of the objectives Should the optics require cleaning only use the finest grade of soft camera lens tissue If foreign matter or stubborn dirt has adhered to a lens surface a small amount of xylene applied sparingly can be applied with ...

Page 12: ...tor Variable 6 volt 20 watt Koehler type vertical halogen illuminator with field diaphragm for both AC and DC operation Filters Green and yellow filters are standard A filter slot is built into the illuminator housing Dimensions Weight Physical weight dimensions 9 lbs 9 x 11 x 10 WDH Shipping weight and dimensions 28 lbs 21 x 14 x 16 Warranty 5 year limited warranty instruction manual and dust cov...

Page 13: ...defects in material and workmanship for a period of five 5 years for mechanical and optical components and one 1 year for electrical components from the date of invoice to the original end user purchaser This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then UNITRON approved service personnel This warrant...

Reviews: