IM CW140-E
3-12
3.6 Connecting a Power Supply and Turning It On/Off
●
Procedure for Connecting the AC Adapter
Follow the steps below to connect the AC adapter.
1. Make sure the power switch on the CW140 is turned off.
2. Attach the clamp filter supplied together with the AC adapter to the output-
side cable of the power cord.
SEE ALSO
The figure given below.
Approx.
10 cm
Side for connection
to the CW140
part number: A1193MN
clamp filter
AC-input side
(for connection to the AC power outlet)
3. Connect the AC adapter plug to the AC adapter jack of the CW140.
4. Connect the plug of the power cord supplied together with the AC adapter
to the power supply connector of the AC adapter.
5. Connect the other end of the power cord to a power outlet that satisfies the
power ratings (requirements) shown below.
Power Ratings of AC Adapter
Supply voltage rating
100 to 240 V AC
Allowable supply voltage range
90 to 264 V AC
Power supply frequency rating
50/60 Hz
Allowable power supply frequency range
48 to 62 Hz
Maximum power consumption
70 to 90 VA
Output voltage rating of AC adapter
12 V DC
Maximum output current rating of AC adapter 2.6 A