IM CW140-E
19-8
(14) Harmonics mode
System
PLL synchronization
Measurement frequency range
Fundamental wave frequency 45
≤
f
≤
65 Hz
Number of analysis orders
1 to 13
FFT data length
512
FFT processing word length
32 bits
Window function
Rectangular
Sampling rate
f
×
256 Hz
Window width
2 periods of f
Display fields
Table display
Voltage and current
RMS, content, phase angle, All-RMS, total
harmonic distortion (IEEE/CSA), fundamental
wave frequency
Power
Power, power content, power phase angle, Total
electric energy, Total power factor, fundamental
wave frequency
Graph display
Voltage and current
Total RMS, content, phase angle
Power
Power, power content, power phase angle
Display accuracy
RMS, power
±(1.5% rdg + 1.5% rng): <1>
Content
Value calculated from <1> ± 2 dgt
Phase angle
± 5 deg
Total harmonic distortion
Value calculated from <1> ± 2 dgt
Logging function
The logging function can be used to take continuous measurements.
START setting
MANUAL, TIME, TRIGGER (controlled)
STOP setting
MANUAL, TIMER, TIME, TRIGGER
Output interval
Can be set from a range of 2 minutes to 1000
hours (in one-minute increments).
The harmonic measure mode does not work with two-load systems.
19.1 Specifications
Summary of Contents for CW140
Page 1: ...User s Manual CW140 CLAMP ON POWER METER IM CW140 E IM CW140 E 3rd Edition July 2001 YK ...
Page 2: ......
Page 14: ......
Page 72: ......
Page 102: ......
Page 138: ......
Page 152: ......
Page 160: ......
Page 200: ......
Page 256: ......
Page 274: ......
Page 276: ......
Page 286: ......
Page 289: ......