IM CW140-E
14-8
14.2 External Control Input
The CW140 can be controlled by an external signal of 5/0 V (high/low (H/L)
levels) to start or stop continuous measurement. Refer to the table below.
This function is available when the external control trigger is selected in each
measurement mode.
Instant Measure mode
Logging
Electric Energy Measure mode
Integration
Demand Measure mode
Demand measurement
Harmonics Measure mode
Logging
●
Input Terminals (START/STOP)
For details on connection, refer to "• Connection of External I/O Terminals" in
Section 2.3.
●
Input Circuit
+5 V
4.7 k
Ω
100 pF
Press the
key to set the CW140 ready for continuous measurement.
5 V
(Open)
Continuous measurement
0 V
(Close)
100
Ω
H
L
Circuit diagram
Start
Stop
Start
Stop
●
Input Levels
Low:
0 to 0.8 V
High: 2.0 to 5.0 V
CUTION
• The allowable voltage range of the input terminals is specified as –0.5 V to 5.5 V. Do not apply
a voltage outside this range, as doing so may damage the input circuit.
• If the CW140 is equipped with an optional D/A output, make sure that the terminals are cor-
rectly connected.