Ammonia HR
1.0 - 50.0
mg/l (N)
Reagent-Kit
Internal Name: 155-AmmoniaHR-pre
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Ammonia HR Kit (PL155-Kit)
Measurement procedure:
1
Prepare 2 Ammonia HR cuvettes (16 mm). Label one as a ZERO cuvette.
2
Open the first cuvette (ZERO cuvette).
3
Fill 0.1 ml distilled water into the cuvette, using a pipette.
4
Open the second vial (sample vial).
5
Fill 0.1 ml sample water in the cuvette.
6
Add 1 x Am. Silic. F5 powder pillow into both cuvettes.
7
Add 1 x Am. Cyan. F5 powder pillow into both cuvettes.
8
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
9
The reagents should now react.
10
Tap TEST to start a 20:00 minute(s) countdown.
11
Place the 16 mm adapter in the PrimeLab.
12
Place the ZERO cuvette in the PrimeLab.
13
Put on the lightshield.
14
Start ZERO measurement.
15
Remove the cuvette again.
16
Insert the sample cuvette in the PrimeLab.
17
Put on the lightshield.
18
Tap TEST to perform the measurement.
Notes:
The test result can be converted into the following unit(s): mg/l NH₃, mg/l NH₄⁺.
Deviations up to 25 % can occur in the low measuring range (0 - 5 mg/l). If you intend to
measure low levels of ammonia, use "ID02 - Ammonia LR".
Adjust strong alkaline or acidic water samples to pH 7 by using 1 mol/l hydrochloric acid or 1
mol/l sodium hydroxide, respectively.
In presence of chlorine, add one drop 0.1 mol/l sodium thiolufate per 0.3 mg/l Cl₂.
T8/204
T8/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...