![Water I.D. Primelab 2.0 Manual Download Page 138](http://html2.mh-extra.com/html/water-i-d/primelab-2-0/primelab-2-0_manual_3564857138.webp)
(23)
Hydrazine
5 - 600
µg/l (N₂H₄)
Liquid
Internal Name: 23-Hydrazine-liq
10ml
65ml PL Hydrazine N°1 (PL65Hydraz1)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 10 drop(s) of PL Hydrazine 1 into the cuvette.
8
Screw the lid back on the cuvette.
9
Gently swirl the cuvette to mix the liquid well.
10
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
11
Tap TEST to start a 05:00 minute(s) countdown.
12
After the lapse of a 05:00 minute(s) countdown the measured value is displayed.
Notes:
The temperature of the water sample should be between 17 °C and 25 °C.
Interferences: i) Hazy test sample, ii) The presence of excessive detergents
T83/204
T83/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...