(160)
Hydrocarbons
0 - 1
NTU (Turb)
Internal Name: 160-Hydrocarbons
10ml
Prime L ab
Measurement procedure:
1
Fill 7.5 ml distilled water in a clean 24 mm cuvette.
2
Fill 2.5 ml sample water in the same cuvette.
3
Screw the lid back on the cuvette.
4
Gently swirl the cuvette to mix the liquid well.
5
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
6
Tap TEST to perform the measurement.
Notes:
Interferences: i) Hazy test sample, ii) The presence of excessive detergents
The result is interpreted as follows: "0" = PASSED, "OR" = FAILED
T84/204
T84/204