(152)
Nitrogen-total HR
(all Steps)
5 - 150
mg/l (N)
Reagent-Kit
Internal Name: 152-NitroTotHR-pre
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
Nitrogen HR Kit (PL152-Kit)
Measurement procedure:
1
Prepare 2 Hydroxide HR cuvettes (16 mm). Label one as a ZERO cuvette.
2
Add 1 x Persulfate powder pillow into both cuvettes.
3
Open the first cuvette (ZERO cuvette).
4
Fill 0.5 ml distilled water in the cuvette.
5
Open the second vial (sample vial).
6
Fill 0.5 ml sample water in the cuvette.
7
Screw the lid back on the cuvette IMMEDIATELY.
8
Shake the vials vigorously for 00:30 minute(s).
9
Tap TEST to start a 00:30 minute(s) countdown.
10
Place cuvettes for 30 minutes at 100 °C in the preheated thermoreactor.
11
Tap TEST to start a 30:00 minute(s) countdown.
12
CAUTION: Cuvettes are hot!
13
Remove the cuvettes from the thermoreactor.
14
Let the cuvettes cool down to at least 25 °C.
15
Add 1 x Reagent A powder pillow into both cuvettes.
16
Screw the lid back on the cuvette IMMEDIATELY.
17
Shake the vials vigorously for 00:20 minute(s).
18
Tap TEST to start a 00:20 minute(s) countdown.
19
The reagents should now react.
20
Tap TEST to start a 03:00 minute(s) countdown.
21
Add 1 x Reagent B powder pillow into both cuvettes.
22
Screw the lid back on the cuvette IMMEDIATELY.
23
Shake the vials vigorously for 00:20 minute(s).
24
Tap TEST to start a 00:20 minute(s) countdown.
25
The reagents should now react.
26
Tap TEST to start a 02:00 minute(s) countdown.
T136/204
T136/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...