(161)
Manganese VLR
0.000 - 0.030
mg/l (Mn)
Tablet
Internal Name: 161-Manganese-VLR
10ml
Manganese VLR N°1 Photometer
(TbsHMagVLR1)
Manganese VLR N°2 Photometer
(TbsPMagVLR2)
Measurement procedure:
1
Fill 10 ml test water into a clean 24 mm cuvette.
2
Screw the lid back on the cuvette.
3
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
4
Start ZERO measurement.
5
Remove the cuvette again.
6
Unscrew the lid from the cuvette.
7
Add 1 Manganese VLR N°1 tablet(s) to the test water in the cuvette.
8
Crush the tablet with a clean stirring rod.
9
Add 1 Manganese VLR N°2 tablet(s) to the test water in the cuvette.
10
Crush the tablet with a clean stirring rod.
11
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
15
Tap TEST to start a 20:00 minute(s) countdown.
16
After the lapse of a 20:00 minute(s) countdown the measured value is displayed.
Notes:
Colour formation is extremely temperature sensitive. A temperature of 20 °C +/- 1 °C gives the
optimum test results.
For optimum test results, the sample needs a standing period of 20 min +/- 1 minute. Further
colour change and colour development after this time should be ignored.
T116/204
T116/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...