(80)
COD MR
0 - 1500
mg/l (O₂)
Reagent-Kit
Internal Name: 80-COD-MR-pre
40
40
60
80
100
40
30
30
20
20
20
20
10
10
0
0
°C
°F
COD MR (tubetest) (COD-80-MR)
Measurement procedure:
1
Prepare 2 COD-MR cuvettes (16 mm). Label one as a ZERO cuvette.
2
Open the first cuvette (ZERO cuvette).
3
Fill 2 ml distilled water into the cuvette.
4
Open the second vial (sample vial).
5
Fill 2 ml sample water in the cuvette.
6
Screw the lid back on both cuvettes.
7
Gently swirl the cuvette to mix the liquid well. Caution, heat development!
8
Place cuvettes for 120 minutes at 150 °C in the preheated thermoreactor.
9
CAUTION: Cuvettes are hot!
10
Remove the cuvettes from the thermoreactor.
11
Let the cuvettes cool down to at least 60 °C.
12
Flip both 16 mm cuvettes to mix the liquid well.
13
Let the cuvettes cool down to room temperature.
14
Place the 16 mm adapter in the PrimeLab.
15
Place the ZERO cuvette in the PrimeLab.
16
Put on the lightshield.
17
Start ZERO measurement.
18
Remove the lightshield.
19
Remove the cuvette again.
20
Insert the sample cuvette in the PrimeLab.
21
Put on the lightshield.
22
Tap TEST to perform the measurement.
23
The measured value is immediately displayed.
Notes:
T57/204
T57/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...