• The adapter to enter 24mm vials is installed properly.
Tap on 'PrimeLab' to start the PrimeLab calibration. Follow the instructions displayed on the
PrimeLab 2.0 screen. Once completed, a message 'Calibration successful' will appear.
In case your PrimeLab is linked to the LabCOM® cloud, a calibration certificate (PDF) will be
available in your account under www.calibrations.labcom.cloud.
• The transparent part of the PrimeLab 2.0 measurement chamber is perfectly clean.
• The light shield is properly set on top of the measurement chamber.
• There is no vial inside the measurement chamber.
A step-by-step procedure will be displayed on your PrimeLab 2.0 screen. Please make sure, that:
PrimeLab calibration
Please perform a PrimeLab calibration prior to NTU-Turbidity calibration.
Use caution to shaking-/rest-instructions on the standard vials
(0.5/10/1000 NTU):
Attention:
• There is no vial inside the measurement chamber.
A step-by-step procedure will be displayed on your PrimeLab 2.0 screen. Please make sure, that:
• The transparent part of the PrimeLab 2.0 measurement chamber is perfectly clean.
• Calibration solution vials are 100% clean, without fingerprints, scratches, spots.
Please choose a cuvette that you use exclusively for the turbidity measurement. It must not be
used for any other test.
• You have properly sealed 24mm glass vials with calibration solutions (not expired) 0.5/10/1000
NTU‘ in hand.
• You always align the arrow on the vial with the arrow on the measurement chamber.
In the "Settings" menu, go to the "Calibration and Indexing" option and then to 'NTU-Turbidity' to
start the calibration process. Follow the instructions displayed on the PrimeLab 2.0 screen.
• The adapter to enter 24mm vials is installed properly.
NTU-Turbidity calibration
Settings
Continued...
22
PrimeLab
2.0
⚙
PTSA calibration
• The 24mm cuvette adapter is correctly inserted,
Please perform a PrimeLab calibration prior to the PTSA calibration.
A step-by-step procedure is displayed on the screen of your PrimeLab 2.0. Please make sure that:
• The measuring chamber of the PrimeLab 2.0 is clean,
• There is no cuvette in the measuring chamber,
• The properly sealed 24 mm cuvettes with calibration solutions (not expired) 0/ 100/400 ppb PTSA
are ready.
• Always align the arrow on the cuvette with the arrow on the measuring chamber.
• All cuvettes are 100% clean, without fingerprints, scratches or stains.
In the "Settings" menu, go to the "Calibration and Indexing" option and then to "PTSA" to start the
calibration process. Follow the instructions displayed on the PrimeLab 2.0 screen.
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...