(97)
Nitrite HR
0 - 1500
mg/l (NaNO₂)
Tablet
Internal Name: 97-Nitrite-HR-tab
10ml
Nitrite HR N°1 Photometer (TbsHNiHR1)
Nitrite HR N°2 Photometer (TbsPNiHR2)
Measurement procedure:
1
Fill 9 ml distilled water in a clean 24 mm cuvette.
2
Fill 1 ml sample water in the same cuvette.
3
Screw the lid back on the cuvette.
4
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
5
Start ZERO measurement.
6
Remove the cuvette again.
7
Unscrew the lid from the cuvette.
8
Add 1 Nitrite HR N°1 Photometer tablet(s) to the test water in the cuvette.
9
Crush the tablet with a clean stirring rod.
10
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
11
Add 1 Nitrite HR N°2 Photometer tablet(s) to the test water in the cuvette.
12
Crush the tablet with a clean stirring rod.
13
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
14
Screw the lid back on the cuvette.
15
Gently swirl the cuvette to mix the liquid well.
16
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
17
Tap TEST to start a 05:00 minute(s) countdown.
18
After the lapse of a 05:00 minute(s) countdown the measured value is displayed.
Notes:
Filter sample if neccessary to test a clear sample.
Make sure the temperature of the sample does not exceed 30 °C.
Chlorine levels above 30 mg/l interfere with this test and can influence the reading.
The test needs to be performed without a delay. Place the vial into the PrimeLab right after
reagents have dissolved and lid is closed. Immediately press TEST. It is essential for the
accuracy of this test to keep the countdown of 05:00 minutes right after dissolving the tablets/
closing the lid/placing the vial into the PrimeLab.
T130/204
T130/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...