(110)
Phosphonate
0.0 - 20.0
mg/l (PO₄³⁻)
Tablet
Internal Name: 110-Phosphon-tab
10ml
Oxidising OP Photometer
(TbsHOXOP)
OP-A Photometer (TbsPOPA)
OP-B Photometer (TbsPOPB)
OP-AX Photometer (TbsHOPAX)
Measurement procedure:
1
Fill 8 ml of distilled water into a clean 24 mm cuvette.
2
Add exactly 2 ml sample water to the same cuvette.
3
Screw the lid back on the cuvette.
4
Swivel the cuvette back and forth for 5 times.
5
Insert the cuvette into the PrimeLab. Pay attention to the arrow on the front of the cuvette.
6
Start ZERO measurement.
7
Remove the cuvette again.
8
Unscrew the lid from the cuvette.
9
Add 1 OrgaPhos-OX tablet(s) to the test water in the cuvette.
10
Crush the tablet with a clean stirring rod.
11
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
12
Screw the lid back on the cuvette.
13
Gently swirl the cuvette to mix the liquid well.
14
Tap TEST to start a 05:00 minute(s) countdown.
15
The cuvette does not need to be placed in the device during this time.
16
Remove the cuvette again.
17
Unscrew the lid from the cuvette.
18
Add 1 OrgaPhos No.1 tablet(s) to the test water in the cuvette.
19
Crush the tablet with a clean stirring rod.
20
Stir with the stirring rod for about 20 seconds until the reagent is completely dissolved.
21
Screw the lid back on the cuvette.
22
Gently swirl the cuvette to mix the liquid well.
23
Tap TEST to start a 02:00 minute(s) countdown.
24
The cuvette does not need to be placed in the device during this time.
25
Remove the cuvette again.
26
Unscrew the lid from the cuvette.
T166/204
T166/204
Summary of Contents for Primelab 2.0
Page 17: ...EMPTY due to technical reasons 16 PrimeLab2 0...
Page 31: ...EMPTY due to technical reasons 30 PrimeLab2 0...
Page 35: ...EMPTY due to technical reasons 34 PrimeLab2 0...
Page 37: ...EMPTY due to technical reasons 36 PrimeLab2 0...
Page 41: ...40 EMPTY due to technical reasons PrimeLab2 0...
Page 45: ...due to technical reasons EMPTY 44 PrimeLab2 0...
Page 252: ...NTU standards should be stored at 5 25 C T197 204 T197 204...
Page 261: ...EMPTY due to technical reasons 48 PrimeLab2 0...
Page 267: ...EMPTY due to technical reasons 54 PrimeLab2 0...
Page 283: ...70 due to technical reasons EMPTY PrimeLab2 0...
Page 293: ...due to technical reasons EMPTY 80 PrimeLab2 0...